Calibrating the NI 449x with DAQmx 8.7

Primary Software: Driver Software>>NI-DAQmx
Primary Software Version: 8.7
Primary Software Fixed Version: N/A
Secondary Software: N/A

Problem:

I am using NI-DAQmx 8.7. When I try to self-calibrate my NI 449x device from MAX, I get an "Internal Software Error." When I try to calibrate the board using LabVIEW or Calibration Executive, I receive Internal Software Error -223043.



Solution:

An issue was identified with the NI 449x that causes these errors when trying to calibrate or self-calibrate the device. The attached zip files (see below) include a patch for DAQmx 8.7 and will address the issue. This issue is limited to DAQmx 8.7 and will be fixed in DAQmx 8.7.1 or later.

Installation Instructions:

  • Download the attached zip file DAQmx87Patch.zip
  • Unzip the file and run the "dsai.msi" installer
  • If you are using a 64-bit windows operating system, you must also download and unzip DAQmx87Patch(64-bit).zip and run the "dsai64.msi" installer.
  • After installation completes, reboot your system.



Related Links:

Attachments:


DAQmx87Patch.zipDAQmx87Patch(64-bit).zip


Report Date: 02/05/2008
Last Updated: 03/10/2008
Document ID: 4I4AOP7E

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit