Error Message "Calibration Sequence Error" when running the Calibration Procedure. Primary Software: Calibration ExecutivePrimary Software Version: 1.0 Primary Software Fixed Version: N/A Secondary Software: N/A
Problem: When I run the Calibration Executive procedure, the first two steps including the self-calibration are properly passed, but the third step is failing. I obtain the following message "Calibration Sequence Error", and the execution stops and a blank report is generated (see below). It occurs in automatic and manual mode. Solution: This error occurs if your computer is set to use a comma instead of a period for decimal points.
Related Links: KnowledgeBase 229ENA8A: Error -15 from E Series Procedure of Calibration Executive Attachments: