Disabling Calibration Verification Using Calibration Executive

Primary Software: Calibration Executive
Primary Software Version: 2.0
Primary Software Fixed Version: N/A
Secondary Software: N/A

Problem: Can I make Calibration Executive verify only once if I disable the "Adjust" step?

Solution: This depends upon the device that is being calibrated. Most devices (E-series, DMM's, etc.) have two different sets of limits corresponding to "Before" calibration and "After" calibration. For these devices, Calibration Executive will verify compliance with both sets of limits, even if the "Adjust" step is disabled. In other words, the board will be verified twice, once for each set of limits. For the devices that only have one set of limits, the "Verify" step will only be executed once if the "Adjust" step is disabled.