Why Isn't My SCXI-1163 Detected in MAX?

Hardware: SCXI>>Modules>>SCXI-1163

Problem:
I have a SCXI-1163 in my SCXI chassis, but when I click the Auto-Detect All Modules button in my SCXI Chassis Configuration dialog box, MAX does not find my SCXI-1163.

Solution:
Depending on the revision of your SCXI-1163, it may or may not be used as a communicating module for a SCXI chassis. Revision E and later of the SCXI-1163 is able to be used as a communicating module in a chassis. To know what revision you have, check the seventh character in the part number. For example, part number 182247C-01 is revision C. If you have a revision earlier than E, then you will need another SCXI module as the communicating module for the chassis.

Related Links:
Products and Services: SCXI-1163

Attachments:





Report Date: 04/06/2007
Last Updated: 04/06/2007
Document ID: 4857D34G

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 