After Power Cycling the DAQPad-1200 with LabVIEW Open, I Get Bad Data when I Run a VI Hardware: Legacy DAQ Devices>>Multifunction DAQ>>DAQPad-1200
Problem: If I power cycle my DAQPad-1200 while LabVIEW is open, then I get bad data when I subsequently run my VI. I have to exit LabVIEW and re-run the configuration utility in order to reset the parallel port. Then I can reopen LabVIEW and get good data. How can I avoid this? Solution: Call the Device Reset.vi at the beginning of your program. Related Links: Product Manuals: DAQPad-1200 User Manual DAQ Support: Data Acquisition Installation Wizard KnowledgeBase 28N9A7QO: Data Acquisition: Troubleshooting Bad Readings Attachments:
Report Date: 06/02/1997 Last Updated: 12/28/2001 Document ID: 0Y1D6AVK |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||