Error -10635 when Running SCXI-1520 Calibration Executive Procedure with SCXI-1001 Chassis Primary Software: Utility Software>>Measurement & Automation ExplorerPrimary Software Version: 3.0 Primary Software Fixed Version: N/A Secondary Software: Calibration Executive Hardware: SCXI>>Chassis>>SCXI-1000, SCXI>>Chassis>>SCXI-1001, SCXI>>Modules>>SCXI-1520
Problem: The following error occurs when running the SCXI-1520 Calibration Executive Procedure with an SCXI-1001 Chassis. Error -10635, Error Message: invalid SCXIOpModeError: Either the SCXI operating mode specified in a configuration call is invalid, or a module is in the wrong operating mode to execute the function call The function SCXI_Set_Config throws an error when using the SCXI-1001 chassis. If you use an SCXI-1000 chassis, you won't see this error. Solution: Configure the SCXI-1001 as an SCXI-1000 in MAX and use only the first four slots of the chassis for your calibration procedure. Related Links: NI SCXI-1102 Product Manuals Attachments:
Report Date: 09/03/2004 Last Updated: 01/17/2005 Document ID: 3D1E7G1W |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||