Thermocouple Measurement Configuration is Incorrect for the SCC-TC0x Modules in VirtualBench-DMM 2.1.2

Primary Software: Other NI Software>>VirtualBench
Primary Software Version: 2.1
Primary Software Fixed Version: N/A
Secondary Software: N/A

Problem: To configure a thermocouple temperature measurement reading in the VirtualBench-DMM a user must specify the CJC source and CJC channel.

When using the SCC-TC0x modules the CJC signal comes in on channel x+8. When this channel is selected in the configuration the measured value is incorrect. BUT, if the user selects "cjtemp" the value is correct. Based on the documentation for the SCC-TC0x module there is no "cjtemp."


Solution: In VirtualBench 2.1.2 this behavior is present, it has been fixed in VirtualBench-DMM 2.51 After upgrading, you can go to Settings under the Edit pull-down menu. Then you will need to set the Temperature Sensor as needed, Temperature Units as needed, TC Cold Junction Source as Thermistor [Vref), and Cold Junction Channel as 8 more than the channel you are reading to get the correct voltage. Using "cjtemp" as you channel will give a realistic value but one that is a little lower than the actual temperature. This should solve any problems with the CJC issue.

Related Links: Products & Services: VirtualBench

Attachments:





Report Date: 04/28/1999
Last Updated: 12/08/2004
Document ID: 1KREGRKS

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit