VirtualBench-Scope 2.1.1 and 2.1.2 Don't Recognize the PXI-5102 or PCI-5102 with NI-DAQ 6.5 or HigherPrimary Software: Other NI Software>>VirtualBenchPrimary Software Version: 2.5 Primary Software Fixed Version: N/A Secondary Software: N/A
Problem: VirtualBench-Scope 2.1.1 and 2.1.2 don't recognize the PXI-5102 or PCI-5102 with NI-DAQ 6.5 or higher. Solution: Starting in NI-DAQ 6.5, the PXI-5102 and PCI-5102 are recognized by the system as "NI 5102" instead of by their former names of "PCI-5102" or "PXI-5102". Since VirtualBench 2.1.1 and 2.1.2 were written before NI-DAQ 6.5, they don't search for "NI 5102"; therefore, they don't recognize the cards. VirtualBench 2.5 (or higher) recognizes the new format. The workaround is to upgrade VirtualBench or downgrade NI-DAQ to 6.1.1 Related Links: VirtualBench Upgrade Information Attachments:
Report Date: 12/01/1998 Last Updated: 12/05/2004 Document ID: 1G09N8TL |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||