Error -10007 in NI-DAQ When Specifying "1,0" as Channel String When Setting the Integrated Circuit Piezolelectric Excitation for a DSA BoardPrimary Software: Driver Software>>NI-DAQPrimary Software Version: 6.1 Primary Software Fixed Version: N/A Secondary Software: LabVIEW Development Systems Hardware: Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PXI-4472B, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4454, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PXI-4474, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4552, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PXI-4461, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PXI-4551, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PXI-4552, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4472, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4551, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4451, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4501, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>NI 4472 for IEEE 1394, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PXI-4471, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4474, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PXI-4462, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4453, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4471, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4452, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PXI-4472
Problem: I get an error -10007 in NI-DAQ when I specify "1,0" as a channel string in LabVIEW using my Dynamic Signal Analyzer (DSA). Is channel order important for the NI 44xx or NI 45xx when programmed in NI-DAQ mode? Solution: With NI-DAQ 6.1 any channel, except "-1", can be specified for the DSA instruments, but the sequence has to be in ascending order. In other words, any channel may appear in the scan list, but it must be in ascending numerical order. Examples of valid scan lists are "0,1" and "0, 2, 3". The DAQ Hardware Overview Guide (DSA chapter - Data Acquisition section, 1st paragraph) explains the proper channel addressing. Note: The channel list to AI Config still must be in ascending order. Related Links: DAQ Hardware Overview Guide KnowledgeBase 1FNCJR0O: Problems with AI_Change_Parameter on DSA Boards Attachments:
Report Date: 03/27/1998 Last Updated: 06/18/2004 Document ID: 17QA876P |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||