Academic Company Events Community Support Solutions Products & Services Contact NI MyNI
16 ratings:
 4.25 out of 5   

Reasons to Ground Thermocouples



Hardware: PXI/CompactPCI>>Controllers

Problem:

Why do I need to ground my thermocouples when I am already measuring them in differential mode?


Solution:

There are two issues involved here:
  1. Static Common Mode Voltage on the Thermocouple Leads:
    Your thermocouple leads can develop a substantial potential difference through charge build-up, which may approach or exceed the maximum common mode voltage rejection of your op amp. This can come from the surrounding environment or from the amplifier/multiplexer circuit itself. In either case, grounding the negative thermocouple lead allows this static charge to discharge to ground and thus preserve the common mode voltage at zero. To ground the signal, place a resistor between the thermocouple lead and the analog ground of the card you are using. An ideal range for these resistors is between 10 kOhms and 100 kOhms. Please, refer the tutorial: Field Wiring and Noise Considerations for Analog Signals.

  2. Injected Current from the Amplifier/Multiplexer Circuit:
    Every op amp generates a Reverse Bias Leakage Current, which causes charge to build up on its positive and negative input leads. This is typically negligible, but it can build up over time, leading to effect listed above. Also, as the multiplexer switches between pairs of input leads, a short-time current (charge injection) is introduced into the new pair of leads. Since thermocouple voltages are typically about 1 mV, these short-time fluctuations can measurably affect the input signal. Again, the solution is to connect the negative thermocouple lead to ground, allowing the injected charge to discharge to ground and not appear at the positive thermocouple lead.


Related Links:
White Paper: Field Wiring and Noise Considerations for Analog Signals
KnowledgeBase 3DM9MEJZ: Thermocouple Reads a Large Negative Value
KnowledgeBase 2Y57ST7Y: Open-Thermocouple Detection
KnowledgeBase 0Z63B4AQ: How Can I Address and Use Cold Junction Compensation (CJC) with SCXI Modules?

Attachments:





Report Date: 09/30/1997
Last Updated: 09/29/2017
Document ID: 11TARTWV

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit