Thermocouple Accuracy Table by Type and Temperature Hardware: Compact FieldPoint, CompactDAQ, CompactRIO, Multifunction DAQ (MIO), SCXI, SCC
Problem: I am reading the temperature of a chamber using two thermocouple of the same type but the readings are different. I have calculated the accuracy of my measurement hardware but it does not compensate for the difference in readings. What may be causing this phenomenon? Solution: In addition to considering the accuracy of the measurement hardware, you should consider the inherent accuracy of the thermocouple. The accuracy of a thermocouple is effected by its type and temperature. See the table below for the maximum permitted error (in degree Celsius) for various common thermocouples that comply with IEC 584-2 (1982).
Related Links: International Electrotechnical Commission (IEC) - Standards organization responsible for thermocouple accurcay standard IEC 584-2. Attachments:
Report Date: 02/06/2007 Last Updated: 03/14/2008 Document ID: 465HI1BH |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||