Academic Company Events NI Developer Zone Support Solutions Products & Services Contact NI MyNI
1 rating:
 1 out of 5     Rate this Document

Using the NI USB-DAQ 9237 with SignalExpress 1.x

Primary Software: LabVIEW Signal Express>>Development System
Primary Software Version: 1.1.1
Primary Software Fixed Version: N/A
Secondary Software: N/A
Hardware: CompactRIO>>Analog Input Modules>>NI 9237

Problem:
I have an NI-USB 9237 module that works fine in Measurement and Automation Explorer (MAX), but it doesn’t show up in the device list when I create a "DAQmx Acquire" step in SignalExpress 1.x, why is this?

Solution:
When you create a DAQmx Acquire step in SignalExpress 1.x, a Voltage task is created. You cannot create a Voltage task with the NI 9237, only Strain and Custom Voltage with Excitation tasks can be created.

A great way to solve this problem is to upgrade to LabVIEW SignalExpress 2.0 as it has the DAQ Assistant integrated into it. LabVIEW SignalExpress 2.0 is included with NI-DAQmx version 8.5 and later. A free download of NI-DAQmx 8.5 is linked below.

If upgrading to LabVIEW SignalExpress 2.0 is not an option, then you can use a LabVIEW Plug-in to access your 9237 with SignalExpress 1.x. Please refer to the linked Developer Zone Tutorial for directions.

Related Links:
Support: NI-DAQmx Version 8.5 for Windows 2000/Vista x64/Vista x86/XP
Developer Zone Tutorial: Creating SignalExpress Plug-Ins with LabVIEW
KnowledgeBase 3WPK09RX: Signal Express User-Defined Steps with LabVIEW 8.x VIs


Attachments:





Report Date: 11/12/2006
Last Updated: 05/08/2007
Document ID: 43BFN44G

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit