Initial High Frequency Waveform When Using the NI ELVIS Function Generator Express VI Primary Software: Driver Software>>NI ELVIS Driver VIsPrimary Software Version: 2.1 Primary Software Fixed Version: N/A Secondary Software: N/A Hardware: Multifunction DAQ (MIO)>>Connector Blocks>>NI ELVIS
Problem: The Function Generator Express VI initially produces a waveform at a much higher frequency than I specified. Is there some way to avoid this? Solution: Each time the Function Generator Express VI starts, the Function Generator is initialized which produces this high frequency signal. If you want to change the waveform without seeing the high frequency waveform, run the Function Generator in a While Loop with the stop condition of the loop wired to the stop terminal of the Function Generator Express VI, as shown below. Related Links: KnowledgeBase 3X4DF8X6: Writing to a High Digital Line with the ELVIS Digital Writer Express VI Causes Glitch KnowledgeBase 3X49ETX6: Changing the Value of the ELVIS Variable Power Supply Express VI Temporarily Sets Output to Zero Attachments: