Cascading the BNC-2096 with an SCXI Chassis Hardware: SCXI>>Terminal Blocks>>BNC-2096
Problem: I have a BNC-2096 (TEDS) and I want to cascade it with my SCXI Chassis. Is this even possible? Solution: Yes, the BNC-2096 (TEDS) can be cascaded with an SCXI Chassis. Cascading the BNC-2096 requires the use of the SCXI-1346 Shielded Multichassis Adapter. Connect the SCXI-1346 to the back of the SCXI Chassis then cable the 68-pin cable from the To Next Chassis connection to the BNC-2096 (TEDS). The BNC-2096 can now be controlled from the same DAQ board that controls the SCXI Chassis. Note: Make sure that the address for the BNC-2096 (TEDS) differs from the address for the SCXI Chassis. Related Links: KnowledgeBase [37IAS2F4]: How Do I Configure My TEDS Sensor Using the BNC-2096? KnowledgeBase [36B9F7CW]: Creating and Writing TEDS Data Example Program: Writing to TEDS using the BNC-2096 Attachments:
Report Date: 10/04/2005 Last Updated: 09/25/2007 Document ID: 3Q3HLT46 |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||