Is the Ethernet Port of the cRIO 9004 Controller Isolated?

Hardware: CompactRIO>>Controllers>>cRIO-9004

Problem:
Is the Ethernet Port of the cRIO 9004 Controller Isolated?

Solution:
The ethernet port is magnetically isolated from the rest of the controller hardware. This should be part of the compliance specifications available operating instrucutions for the cRIO 9004. If more isolation is needed there are also 3rd-party opto-isolators available, which can help cut down on ground loops and block voltage spikes.

Related Links:
Compact RIO cRIO-9002/9004 Operating Instructions