NI 6123 Offset After Calibration Primary Software: Calibration ExecutivePrimary Software Version: 3.1 Primary Software Fixed Version: N/A Secondary Software: N/A Hardware: Multifunction DAQ (MIO)>>S Series>>PCI-6123, Multifunction DAQ (MIO)>>S Series>>PXI-6123
Problem: When calibrating a NI 6123 with Calibration Executive 3.x, an offset occurs which causes the device to fail calibration. What can I do to prevent this problem? Solution: To address this issue, install NI-DAQmx version 8.1 or later, after you install Calibration Executive. Related Links: NI Calibration Executive Homepage Current DAQmx Driver Download Page Attachments:
Report Date: 05/15/2006 Last Updated: 06/15/2006 Document ID: 3XECTLOR |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||