Setting Up the SCC-ACC01 in a Dual-Stage Configuration Hardware: SCC>>Carriers>>SC-2345
Problem: I have an SCC-ACC01 (formerly SCC-Integrated Circuit Piezoelectric01) and an SCC-LP0X lowpass filter module. I want to setup the SCC-ACC01 in a dual-stage configuration. When I try to configure the SC-2345 in Measurement & Automation Explorer (MAX) I get an error stating that the SCC-ACC01 is not allowed in the first stage of a dual-stage configuration. Solution: The SCC-ACC01 module can be the first stage of a dual-stage configuration. However, this configuration is not supported by MAX 3.0 and earlier. To take a measurement with the SCC-ACC01 in a dual-stage configuration, skip the SC-2345 configuration in MAX. Read the raw voltage from the SCC-ACC01 channel and scale it with the following formula:
Related Links: Product Manuals: SCC-ACC01 Accelerometer Input Module User Guide Attachments: [B. Boldt, 8/8/03] Minor Edits
Report Date: 08/06/2003 Last Updated: 07/21/2004 Document ID: 305DCOYB |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||