Hardware Timed Single Point with S Series Devices Hardware: Multifunction DAQ (MIO)>>S Series
Problem: Is Hardware Timed Single Point available for S Series devices? Solution: Hardware-Timed Single Point is not available for the 611x, 612x, or 613x devices, this limitation comes from the typically deep FIFOs on those devices. Hardware-Timed Single Point is available for the 6143 if you use an even number of channels, and for the 6154 when using any number of channels. Note: The Specifications tab on the product pages for the 611x, 612x, and 613x devices mention Deterministic single-point control, this is referring to the ability to control the sampling rate using the On Demand sampling mode with the Real-Time OS. Related Links: Products and Services: S Series Simultaneous-Sampling Multifunction DAQ Attachments:
Report Date: 05/22/2007 Last Updated: 05/22/2007 Document ID: 49LGK84G |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||