PXI-4072 Failing Calibration in Calibration Executive 3.2.1

Primary Software: Calibration Executive
Primary Software Version: N/A
Primary Software Fixed Version: N/A
Secondary Software: N/A

Problem:
When performing a calibration of a PXI-4072 in Calibration Executive 3.2.1 my calibration always fails. Why does the failure always occur during the Capacitance Verification step?

Solution:
The PXI-4072 procedure had a problem in the Capacitance Verification step in Calibration Executive 3.2.1. This is fixed in Calibration Executive version 3.2.1.f1 and later. You can find the Calibration Executive update link here.

You may also see an error when calibrating if you are using the NI-DMM 2.7.0 driver. See KB 4ALFRQE5 linked below for more details.

Related Links:
KnowledgeBase 4ALFRQE5: Error 1074118637 in Calibration Executive when Calibrating a DMM
Drivers and Updates: Calibration Executive 3.2 - Windows - 3.2 Updates
Calibration Resources: Calibration Solutions from National Instruments