Low Voltage and Current Measurements Are Noisy when Using NI ELVISPrimary Software: Driver Software>>NI ELVIS Driver VIsPrimary Software Version: 1.0 Primary Software Fixed Version: N/A Secondary Software: N/A
Problem: I am using the DMM within NI ELVIS to take voltage and current measurements. I have noticed that when taking very small voltage and current measurements, the readings fluctuate and seem noisy. Is this noise inherent to NI ELVIS or is there something I can change in software to decrease the fluctuations in readings? Solution: When using the DMM to take voltage and current measurements, the DMM is only acquiring 100 points to average by default. The number of samples needs to be increased to account for noise in the measurement. The current measurement is greatly affected by the low number of samples being taken and the AC voltage is affected when performing measurements on a signal with a frequency below 100 Hz. Since NI ELVIS gives the user access to the source code, you can modify the DMM source code to increase the sample size when performing voltage and current measurements. The steps for increasing the sample size are listed below.
Related Links: Attachments:
Report Date: 06/24/2003 Last Updated: 09/22/2004 Document ID: 2YN9ST0T |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||