Voltage Signals Merging Together

Hardware: SCXI>>Modules>>SCXI-1100

Problem: Why are my voltage signals merged together when using the SCXI-1100?

Solution: The reason why the signals are merging together is because the filter setting on the module is too high. The filter setting can be changed by switching a jumper on the module. To do this, follow the steps below
  1. If the module is still in the chassis, turn off the chassis and remove the module.
  2. Remove the module's cover.
  3. Remove the jumper from W4 (4 Hz filter) and connect it to W3 (10 kHz filter) or W2 (no filter).
  4. Replace the module's cover.
For detailed information on how to remove the module's cover refer to page 38 of the Getting Started with SCXI Manual (linked below). For detailed information regarding the SCXI-1100 filter settings refer to page 25 of the SCXI-1100 User Manual (linked below).