NI ELVIS Express VIs Reset the FGEN Output When FGEN Express VI is Run in a Separate VIPrimary Software: Driver Software>>NI ELVIS Driver VIsPrimary Software Version: 2.1 Primary Software Fixed Version: 2.1 Secondary Software: N/A Hardware: Multifunction DAQ (MIO)>>Connector Blocks>>NI ELVIS
Problem: If I put the FGEN Express VI in a while loop in one VI and run it, and in another VI run one of the other instrument's (such as the DMM or Scope) Express VI, the second VI doesn't alter its output because the FGEN resources are still reserved by the first VI. Solution: If you run the two Express VIs in the same VI and wire them together using the error cluster (see image), the second Express VI does not reset the FGEN output to the default sine wave.
Related Links: Attachments:
Report Date: 05/05/2006 Last Updated: 05/12/2006 Document ID: 3WA9F3WW |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||