Calibration Executive is giving Error -1074001424 When Calibrating a PCMCIA-4050Primary Software: Calibration ExecutivePrimary Software Version: 3.2 Primary Software Fixed Version: N/A Secondary Software: N/A
Problem: Why am I receiving error -1074001424 trying to calibrate my PCMCIA-4050 with Calibration Executive? Solution: Prior to Calibration Executive 3.2.2, the software never checked to make sure that the correct calibrator was being used. Calibration Executive 3.2.2 will throw error -1074001424 when trying to calibrate a PCMCIA-4050, if the wrong calibrator is being used. Only the Fluke 5700A and 5720A calibrators can be used with the PCMCIA-4050. These calibrators have to be used because of their precision. You can not use a Fluke 5500A or 5520A Calibrator. This information is outlined in the Calibration Help File. (Attached) Related Links: Calibration Executive Help Attachments:
Report Date: 12/27/2007 Last Updated: 12/28/2007 Document ID: 4GQAL51U |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||