Academic Company Events NI Developer Zone Support Solutions Products & Services Contact NI MyNI
1 ratings:
 2 out of 5     Rate this Document

Calibration Executive is giving Error -1074001424 When Calibrating a PCMCIA-4050

Primary Software: Calibration Executive
Primary Software Version: 3.2
Primary Software Fixed Version: N/A
Secondary Software: N/A

Problem:
Why am I receiving error -1074001424 trying to calibrate my PCMCIA-4050 with Calibration Executive?

Solution:
Prior to Calibration Executive 3.2.2, the software never checked to make sure that the correct calibrator was being used.  Calibration Executive 3.2.2 will throw error -1074001424 when trying to calibrate a PCMCIA-4050, if the wrong calibrator is being used. 

Only the Fluke 5700A and 5720A calibrators can be used with the PCMCIA-4050.  These calibrators have to be used because of their precision.  You can not use a Fluke 5500A or 5520A Calibrator.  This information is outlined in the Calibration Help File. (Attached) 

Related Links:
Calibration Executive Help

Attachments:





Report Date: 12/27/2007
Last Updated: 12/28/2007
Document ID: 4GQAL51U

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit