Shared Variable and/or PNG Support Is Not Active After Rebooting the TPC-2006Primary Software: LabVIEW Modules>>PDA ModulePrimary Software Version: 8.0 Primary Software Fixed Version: N/A Secondary Software: LabVIEW Modules>>Touch Panel Module Hardware: Industrial Communications>>HMI and Industrial PC>>TPC-2006
Problem: I installed shared variable and PNG support for my TPC-2006 touch panel and both work. After rebooting, these features stop working and appear to have been uninstalled, and I have to reinstall support every time I reboot to get them working again. How can I force the TPC-2006 to keep shared variable and/or PNG support installed? Solution: The TPC-2006 uses a default image when it reboots, so any support that was installed will be overwritten. This issue only happens when installing the shared variable and PNG support from the LabVIEW PDA module, and not the Touch Panel Module (8.2 or later). There are two ways to force the TPC to keep this support installed:
Note: If the folder \harddisk\Wince\config does not exist, create it. Related Links: Attachments:
Report Date: 11/07/2006 Last Updated: 06/19/2007 Document ID: 4361GBMK |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||