Error 1074118637 in Calibration Executive when Calibrating a DMM Primary Software: Calibration ExecutivePrimary Software Version: 1.0 Primary Software Fixed Version: N/A Secondary Software: Driver Software>>NI-DMM
Problem: The error below is occuring in Calibration Executive when trying to calibrate a DMM from National Instruments: Error -1074118637 occurred at niDMM Read.vi at step Re-verify ACVoltage Possible reason(s): There is an error in the calibration constants. This measurement cannot be made until the board is recalibrated.Solution: This is a known issue with NI-DMM 2.7.0 and Calibration Executive . If this error is received it means that the AC voltage calibration constants on the board are incorrect. Most likely the board passed pre-verification, meaning that the board was in a good state prior to calibration. However, the adjustment stage during calibration did not function properly. After adjustment, Calibration Executive verifies the board again, and at this point the calibration constants are discovered to be incorrect. To correct this issue:
Related Links: Drivers and Updates: NI-DMM Support: Calibration Executive Help Support: Calibration Solutions from National Instruments KnowledgeBase 4926CS17: PXI-4072 Failing Calibration in Calibration Executive 3.2.1 Attachments:
Report Date: 06/22/2007 Last Updated: 05/08/2008 Document ID: 4ALFRQE5 |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||