PXI-4224 Does Not Meet Bandwidth Specifications Hardware: Multifunction DAQ (MIO)>>SC Series>>PXI-4224
Problem: The PXI-4224 manual states that the filter has a cutoff frequency of 30KHz. Why can't I get correct measurements above 15KHz? Solution: Older manuals for the PXI-4224 had incorrectly listed the filter as having a -3 dB cutoff of 30KHz. The correct value is 15KHz at -3dB. This spec has been corrected for the current and future releases of the manual. Related Links: Product Page: NI PXI-4224 Attachments:
Report Date: 08/25/2006 Last Updated: 08/25/2006 Document ID: 40OGANHF |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||