PXI-4224 Does Not Meet Bandwidth Specifications

Hardware: Multifunction DAQ (MIO)>>SC Series>>PXI-4224

Problem:
The PXI-4224 manual states that the filter has a cutoff frequency of 30KHz. Why can't I get correct measurements above 15KHz?

Solution:
Older manuals for the PXI-4224 had incorrectly listed the filter as having a -3 dB cutoff of 30KHz. The correct value is 15KHz at -3dB. This spec has been corrected for the current and future releases of the manual.

Related Links:
Product Page: NI PXI-4224

Attachments:





Report Date: 08/25/2006
Last Updated: 08/25/2006