Error 10401 Running NI 4350/4351 Calibration Procedure in Calibration Executive. Primary Software: Calibration ExecutivePrimary Software Version: 1.0 Primary Software Fixed Version: N/A Secondary Software: Driver Software>>NI-DAQ Hardware: Modular Instruments>>High-Precision DAQ (Dataloggers)>>DAQCard-4350, Modular Instruments>>High-Precision DAQ (Dataloggers)>>DAQPad-4350 for USB, Modular Instruments>>High-Precision DAQ (Dataloggers)>>PXI-4351, Modular Instruments>>High-Precision DAQ (Dataloggers)>>PC-4350, Modular Instruments>>High-Precision DAQ (Dataloggers)>>DAQPad-4351 for IEEE 1394, Modular Instruments>>High-Precision DAQ (Dataloggers)>>PCI-4351
Problem: When running the NI 4350/4351 Calibration Procedure 1.0 with NI-DAQ 6.8.1, I get Error 10401. What is causing this? Solution: There is an update available for the NI 4350/4351 1.0 Calibration Procedure that addresses this issue. This update replaces version 1.0 of the NI 4350/4351 calibration procedure. The original procedure incorrectly adjusted the excitation current on the NI 4350/4351 to 20 uA instead of 25 uA. This may result in Error 10401 due to NI-DAQ 6.8.x and later not recognizing the NI 435x board. This update is available through the link below. Related Links: Technical Support: Calibration Resources Attachments:
Report Date: 10/12/2001 Last Updated: 06/19/2008 Document ID: 2EB8KOB5 |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||