Why do I Get Error -201028 When I import My Measurement and Automation Explorer (MAX) Configuration to a Simulated Device? Primary Software: Utility Software>>Measurement & Automation ExplorerPrimary Software Version: 1.0.1 Primary Software Fixed Version: N/A Secondary Software: N/A Hardware: Multifunction DAQ (MIO)>>E Series
Problem: When I try to export my configuration from one machine and import it to another using a simulated device, I get an import report similar to the following: NI-DAQmx 8.5: For example, if the configuration is made for a 6035E card which is Dev1; on computer 1 and exported to a simulated device on computer 2 which already has a USB-6008 as Dev1; this error will occur. To solve this issue, rename one of the devices which share a name to one which is not in use. In the example, if the USB-6008 on the target computer is renamed to Dev2; and the import is retried, it will be successful. Related Links: KnowledgeBase 2GJJ9SY1: Saving the System Configuration in Measurement & Automation Explorer (MAX) Attachments:
Report Date: 04/11/2007 Last Updated: 05/13/2008 Document ID: 48ABOE3U |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||