Why Won't the SCXI-1100 Sample Faster Than 100k S/s?

Hardware: SCXI>>Modules>>SCXI-1100

Problem:
The SCXI-1100 has a maximum sampling rate of 166k S/s.  However, anytime I set the sampling rate above 100k S/s I get an error.  Why is this happening?

Solution:
Currently the DAQmx driver does not allow the card to sample at a rate that would drop the module's resolution below 14 bits.  In this case this maximum rate for 14-bit resolution comes out to 100k S/s even though the module can go faster.

To achieve the module's actual maximum rate of 166k S/s (and the corresponding 13-bit resolution), manually set the AIConvert.Rate property to 166k S/s with a Timing Property Node in LabVIEW as shown below.



Related Links:
KnowledgeBase 27SDGMCM:  Determining Maximum Scan Rate for Multiple SCXI Modules
KnowledgeBase 2XPE1QCW:  How is the Convert (Channel) Clock Rate Determined in NI-DAQmx and Traditional NI-DAQ?