Measuring DC Voltages with the NI USB-9233 Dynamic Signal Acquisition DeviceHardware: Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>USB-9233
Problem: Can the NI USB-9233 DSA device measure DC voltages? Solution: No. The NI USB-9233 is an AC-coupled device, which means that there is a capacitor in series with the inputs that blocks DC voltages and offset. This is further explained in the Operating Instructions and Specifications Manual linked below. The specifications state that the lower AC cutoff frequency (-3dB point) is at 0.5 Hz. Related Links: Product Manuals: NI 9233 Operating Instructions and Specifications (September 2008) KnowledgeBase 1H59Q2W3: Basic Information about AC and DC Coupling Developer Zone Tutorial : AC and DC Coupling Attachments:
Report Date: 08/29/2005 Last Updated: 10/03/2008 Document ID: 3OS7GQ9N |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||