Academic Company Events NI Developer Zone Support Solutions Products & Services Contact NI MyNI
This Document is not yet Rated  Rate this Document

Scaling in VirtualBench-Logger

Primary Software: Other NI Software>>VirtualBench
Primary Software Version: 2.7
Primary Software Fixed Version: N/A
Secondary Software: N/A

Problem: I would like to apply a scale to my acquired data in VirtualBench-Logger. Is this possible?

Solution: Yes, it is possible to apply a Linear (Mx + B) scale to channels in VirtualBench-Logger. To configure the scaling, select Edit » Settings. Only channels with a Transducer type of VDC, Current, or Resistance can be scaled. When a channel with one of these three types is selected, the Mx + B button can be pressed to select the desired Factor (M), Offset (B), and Engineering Unit parameters for the linear scale. To have the Factor (M) and Offset (B) parameters automatically determined based on Maximum and Minimum unscaled (Raw Scale) and scaled (Engineering Scale) values, you can press the Calculate Scaling button. After a linear scale has been configured for one channel, it can be applied to all channels by pressing the Apply to All Channels button.

To obtain further information concerning this topic, please refer to the VirtualBench-Logger Online Help, available from Help » Online Help.

Related Links: Product Manuals: Getting Started with VirtualBench


Attachments:





Report Date: 10/31/2003
Last Updated: 09/15/2004
Document ID: 32UFF6T0

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit