What Is the Difference Between the 4 Hz Filters on the SCXI-1100, SCXI-1120, and SCXI-1121 Modules?Hardware: SCXI>>Modules>>SCXI-1100
Problem: What is the difference between the 4 Hz filter on the SCXI-1100 module and the SCXI-1120 and SCXI-1121 modules? Solution: On the SCXI-1100 module, the 4 Hz filter is placed after the multiplexer. This lengthens the settling time of the gain amplifier, hence restricting maximum sampling rate to be approximately 2 Hz. On the SCXI-1120 and SCXI-1121, there is a separate 4 Hz filter for each of the input channels, so you can acquire data at a sample rate of 142 kHz. Related Links: Attachments:
Report Date: 01/17/1996 Last Updated: 09/12/2007 Document ID: 0HG6BMRP |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||