Using the Measurement Mode Structure Input to AI Config VI with an E-Series DevicePrimary Software: Driver Software>>NI-DAQPrimary Software Version: 7.2 Primary Software Fixed Version: N/A Secondary Software: LabVIEW Development Systems Hardware: Modular Instruments>>High-Precision DAQ (Dataloggers)>>DAQCard-4350, Modular Instruments>>High-Precision DAQ (Dataloggers)>>DAQPad-4350 for USB, Modular Instruments>>Digital Multimeters (DMM)>>PCI-4060, Modular Instruments>>Digital Multimeters (DMM)>>PXI-4060, Modular Instruments>>High-Precision DAQ (Dataloggers)>>DAQPad-4352 for USB, Modular Instruments>>High-Precision DAQ (Dataloggers)>>PXI-4351, Modular Instruments>>Digital Multimeters (DMM)>>PC-4060, Modular Instruments>>High-Precision DAQ (Dataloggers)>>PC-4350, Modular Instruments>>High-Precision DAQ (Dataloggers)>>DAQPad-4351 for IEEE 1394, Modular Instruments>>High-Precision DAQ (Dataloggers)>>PCI-4351, Multifunction DAQ (MIO)>>E Series
Problem: I noticed that AI Config VI has an input used to specify the measurement mode. The LabVIEW help for AI Config VI does not indicate the valid measurement modes for different devices. I am currently using an E-Series device. Which values for the measurement mode structure input to AI Config VI are valid for E-Series devices? Solution: The measurement mode structure input to AI Config VI is a cluster of two elements: measurement mode and reserved. The only value for measurement mode that is valid for E-Series devices is "no change". Selecting any other value will result in a -10403 error. The measurement mode structure input to AI Config VI is intended to be used with 4060 and 435X devices. Related Links: Attachments:
Report Date: 03/28/2003 Last Updated: 10/12/2006 Document ID: 2VRC1FT0 |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||