Why Can't I Perform a Complete Calibration with Exceptions on an NI 4070/4072 in Calibration Executive?Primary Software: Calibration ExecutivePrimary Software Version: 3.3.1 Primary Software Fixed Version: N/A Secondary Software: N/A
Problem: The NI 4070/4072 6½-Digit FlexDMM Calibration Procedure states that I can perform a complete calibration with exceptions; however, when I uncheck all the boxes for a particular measurement type (AC Voltage for example) the continue button is grayed out. Why is this happening? Solution: At this time it is only possible to skip some steps, not entire measurement types as the documentation states in Appendix A. You may deselect some of the steps to save time as long as one step per measurement type remains selected. Related Links: Product Manual: NI 4070/4072 6½-Digit FlexDMM Calibration Procedure KnowledgeBase 2Q7AN5B9: Calibration Executive Report Specification Limits KnowledgeBase 42H8K3E1: Error 1074135039 With Calibration Executive and PXI-4060 Attachments:
Report Date: 05/15/2009 Last Updated: 05/27/2009 Document ID: 4XEISB28 |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||