Can I Use the 5B37 Thermocouple Module with the DAQ Channel Wizard? Hardware: Signal Conditioning Other>>5B Series
Problem: Can I Use the 5B37 Thermocouple Module with the DAQ Channel Wizard? Solution: Because the 5B37 has a voltage nonlinearity, it cannot be used with the DAQ channel wizard. Rather than using virtual channels, you can use the 5B37 Linearization Library (in the related links, below) to convert the 5B37 voltage value to the corresponding linearized temperature in LabVIEW. For further information, please refer to the page 2-11 of the User Manual and to the Tutorial Note, which are also both linked below. Related Links:
Attachments:
Report Date: 04/15/2002 Last Updated: 08/02/2007 Document ID: 2KEANPVS |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||