Why Do I Get Error -200077 When Trying to Configure My SCXI-1120/D? Hardware: SCXI>>Modules>>SCXI-1120D
Problem: I have a SCXI-1120/D connected to an M-Series card. When I try to acquire data in a Measurement & Automation Explorer (MAX) Test Panel, I receive the following error:
Error -200077 occurred at Test Panel
My M-Series card should be able to allow voltages between -10 and 10 volts. What is going on?Solution: You are experiencing this error because the default gain setting for the SCXI-1120/D is 1000. You can change the gain setting in MAX by right-clicking the module and selecting Properties. In the Jumpers tab, you can change the gain and filter settings for the SCXI-1120/D. If you want to apply no gain, then simply set the gain to 1.
Note: Make sure the gain and filter settings you specify in software match the physical jumper locations on SCXI-1120/D. These jumper locations can be found in the SCXI-1120/D User Manual linked below. Related Links: Product Manuals: SCXI-1120/D User Manual Attachments: