Academic Company Events NI Developer Zone Support Solutions Products & Services Contact NI MyNI
1 ratings:
 5 out of 5     Rate this Document

Error -1074135028 When Closing Session to NI 5102

Hardware: Modular Instruments>>High-Speed Digitizers (Scopes)>>PCI-5102, Modular Instruments>>High-Speed Digitizers (Scopes)>>PXI-5102

Problem:
I am using an NI 5102 digitizer with the NI-SCOPE 3.5.1 driver. Whenever I stop my program using the niScope Close VI, I receive an error -1074135028, as seen below. What is causing this error and how can I fix it?



Solution:
NI-SCOPE 3.5.1 originally contained a problem which causes the NI 5102 to return an Attribute ID not recognized error when the niScope Close VI is called. This has been corrected in the version posted on Drivers and Updates, but if you downloaded the driver between April 1st and April 30th 2009, or installed version 3.5.1 from CD, you may encounter this error.

To solve the problem, you can download the attached niScopeDAQ2u.dll and replace the existing one located in the C:\WINDOWS\system32 directory. If you choose to ignore the error, the device can still be used.

Related Links:
Drivers and Updates: NI-SCOPE 3.5.1 - Windows 2000/Vista x64/Vista x86/XP

Attachments:
niScopeDAQ2u.dllniScopeDAQ2u.dll





Report Date: 05/06/2009
Last Updated: 05/08/2009
Document ID: 4X5GL5LY

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit