我能同步多个RF设备吗?



硬件: Modular Instruments>>RF Measurement Devices>>PXI-5650, Modular Instruments>>RF Measurement Devices>>PXI-5671, Modular Instruments>>RF Measurement Devices>>PXI-5652, Modular Instruments>>RF Measurement Devices>>PXI-5651, Modular Instruments>>RF Measurement Devices>>PXI-5660, Modular Instruments>>RF Measurement Devices>>PXI-5661, Modular Instruments>>RF Measurement Devices>>PXI-5670

问题: 是否可能实现下列RF设备中任意2个或者多个的相位同步,例如NI PXI-565x, NI PXI-5660, NI PXI-5661, NI PXI-5670, or the NI PXI-5671

解答:

简单地回答是不可以。尽管您可以将所有的设备同步到同一个10Mhz的参考时钟,并且同步它们的触发,每个设备都有各自的板上晶振,它们是无法在多个设备间共享的。每个板上晶振都容易受到内在相位漂移和噪声的影响,而这些将影响各个设备的性能。结果就是任何2个有板上晶振的设备都无法精确地同步。

下图显示了一个如何使用板上晶振来混合信号的过程,使之从中频IF混合到高频RF.事实上,NI5600和NI5610使用3种这样的混频器来转换信号。

 

>

注意:所有NI的数字化仪和函数发生器都能相互同步,只要它们没有用在任何上变频或者下变频的过程中。