Error -10007 AI Group Config When Using SCXI-1100 Voltage VI with SCXI-1102Primary Software: LabVIEW Development Systems>>Full Development SystemPrimary Software Version: 7.1 Primary Software Fixed Version: N/A Secondary Software: Driver Software>>NI-DAQ Hardware: SCXI>>Modules>>SCXI-1102, SCXI>>Modules>>SCXI-1100
Problem: When using the SCXI-1100 Voltage VI to test the SCXI-1102, I get an error -10007 AI Group Config, which results in no acquisition. Solution: The SCXI-1100 example is meant for use with only the SCXI-1100. The example attempts to scan the calgnd channel which does not exist on the SCXI-1102 module. You should use the SCXI-1102 thermocouple example which was explicitly made for the SCXI-1102. Related Links: Attachments:
Report Date: 03/04/1996 Last Updated: 06/29/2004 Document ID: 0J3E4JNN |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||