NI-5600 or NI-5610 Calibration FailsPrimary Software: Calibration ExecutivePrimary Software Version: 3.3.1 Primary Software Fixed Version: N/A Secondary Software: N/A
Problem: I use Calibration Executive 3.3.1 to calibrate a PXI-5600 or a PXI-5610 that should pass but the device fails the calibration. Or, I have noticed that the limits in Calibration Executive do not seem to match the specifications. Why is this and how can I fix it? Solution: The device is failing calibration is because the limit data for NI 5600 and NI 5610 are incorrect (i.e. too tight) in Calibration Executive 3.3.1. You can fix this by replacing the old limit file with the new corrected data file which can be found in the attachment section of this knowledgebase. Following is the procedure to install them.
Related Links: Products and Services: NI Calibration Executive KnowledgeBase 2Q7AN5B9: Calibration Executive Report Specification Limits Attachments:
Report Date: 05/20/2009 Last Updated: 06/24/2009 Document ID: 4XJGRGBH |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||