PXI 計測器のアドレスを指定すると、エラー-1074118655、-1073807343、-1074118644 が発生します



ハードウェア: Modular Instruments>>Digital Multimeters (DMM), Digital I/O (DIO)>>High-Speed Digital I/O>>PCI-6562, Modular Instruments>>High-Speed Digitizers (Scopes), Digital I/O (DIO)>>High-Speed Digital I/O>>PCI-6551, Digital I/O (DIO)>>High-Speed Digital I/O>>PXI-6541, Digital I/O (DIO)>>High-Speed Digital I/O>>PXI-6561, Digital I/O (DIO)>>High-Speed Digital I/O>>PXI-6551, Digital I/O (DIO)>>High-Speed Digital I/O>>PCI-6561, Digital I/O (DIO)>>High-Speed Digital I/O>>PXI-6542, Digital I/O (DIO)>>High-Speed Digital I/O>>PXI-6552, Digital I/O (DIO)>>High-Speed Digital I/O>>PXI-6562, Digital I/O (DIO)>>High-Speed Digital I/O>>PCI-6542, Digital I/O (DIO)>>High-Speed Digital I/O>>PCI-6552, Digital I/O (DIO)>>High-Speed Digital I/O>>PCI-6541, Switches, Signal Sources (AWG/FG/AO)>>Arbitrary Waveform and Function Generators

問題:
PXI 計測器(NI-DMM、NI-FGEN、NI-Scope、NI-HSDIO、NI-Switch)のセッションを開くもしくは NI Switch Executive で仮想デバイスを評価しようとしたとき、下記のエラーが発生します。なぜでしょうか? 
 
ドライバ エラーコード

NI-DMM

-1074118655

NI-Switch

-1074118644

NI Switch Executive

-1074118644

NI-FGEN

-1073807343

NI-Scope

-1073807343

NI-HSDIO

-1073807343

 



解決策:
これらのエラーは通常、DAQ リソースデスクリプタ(例、従来型 DAQ では"DAQ::1::INSTR"、DAQmx では"Dev1もしくは PXI1Slot1")の代わりに PXI リソースデスクリプタ(例、PXI1::15::INSTR)で計測器をアドレス指定した際に発生します。NI モジュール式計測器ドライバはボードの検出と通信で NI-DAQ を使用するため、これらのドライバでセッションを開くには DAQ デスクリプタを使用しなければいけません。

NI Switch Executive では、リソースデスクリプタはハードウェアタブの下にある IVI ドライバセクションにあります。 リソースデスクリプタでダブルクリックし、変更してください。

MAX画面

DAQ リソースデスクリプタの詳細については技術データベース 33A96PCW (下記のリンク)を参照してください。



関連リンク:
製品とサービス: モジュール式計測器
技術データベース 33A96PCW: NI-DMMのサンプルVIを実行すると、Error -1074118656が発生します

添付:





報告日時: 10/28/2004
最終更新日: 04/18/2011
ドキュメントID: 3ER6NTR9