Calibration Executive Report Specification Limits Primary Software: Calibration ExecutivePrimary Software Version: 2.0 Primary Software Fixed Version: N/A Secondary Software:
Problem: What do the "Before (As Found)" and "After (As Left)" specification limits correspond to in the Calibration Executive report? Solution: The general rule for these limits is that the "Before" limits correspond to the 1 year error limits, and the "After" limits correspond to the 24 hour error limits associated with that particular board. The 1 year/24 hour limits hold true for devices that have 2 sets of limits. Examples of these include E/S Series, DMMs, Dataloggers (435x), and DSA products. Other products, including oscilloscopes, FieldPoint, and SCXI, only have 1 set of limits. For this situation, the limits are the same for "Before" as compared to "After". Related Links: Attachments:
Report Date: 10/08/2002 Last Updated: 10/09/2002 Document ID: 2Q7AN5B9 |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||