Calibration Executive Report Specification Limits

Primary Software: Calibration Executive
Primary Software Version: 2.0
Primary Software Fixed Version: N/A
Secondary Software:

Problem: What do the "Before (As Found)" and "After (As Left)" specification limits correspond to in the Calibration Executive report?

Solution: The general rule for these limits is that the "Before" limits correspond to the 1 year error limits, and the "After" limits correspond to the 24 hour error limits associated with that particular board. The 1 year/24 hour limits hold true for devices that have 2 sets of limits. Examples of these include E/S Series, DMMs, Dataloggers (435x), and DSA products. Other products, including oscilloscopes, FieldPoint, and SCXI, only have 1 set of limits. For this situation, the limits are the same for "Before" as compared to "After".